Home Products Hot Products New Arrivals News Showroom Service Contact us
  TEL: 86-0755-27192975-801
Search:
  0 item(s) View Cart
Popular Searches:
  Categories
  Reusable SpO2 sensors
  Disposable SpO2 Sensor
  SpO2 adapter Cable
  ECG cable
  ECG Holter Cables
  EKG cable
  ECG /EKG Adapter Cable
  IBP Cable
  Temperature probe
  NIBP Cuff
  original product
  On Sale Information
  SpO2,ecg ,ekg parts
  pluse oximeter
  cable samples
  Ultrasound and Fetal Probe
  EEU Electrode
  Certificate
  Shop With Confidence
  Contact us
86-0755-27192975-801
86-0755-27192975
info@bestspo2.com
wechat:dengyan6572
bestspo2
  News HomeNews
Analysis of Influencing Factors on Pulse Oxygen Saturation Monitoring
Objective To analyze influencing factors on pulse oxygen saturation(SpO2) monitoring for clinical practice reference.Methods The influencing factors of which SpO2 readings were different from arterial oxygen saturation(SaO2) values of 85 cases in central ICU from January 2007 to July 2009 were analyzed retrospectively.Results The abnormal site of SpO2 probe,movement artifact,low perfusion states,abnormity of SpO2 monitoring system,abnormal skin color and ambient light might all lead the inaccurateness of SpO2 measurements.The primary influencing factor was the abnormal site of SpO2 probe,which took a 29.41%(25/85) proportion.Conclusion SpO2 readings can better reflect the changes in SaO2,but many factors influence SpO2 readings.Appropriate intervention is needed to improve its accuracy.
Copyright © 2006-2012 Shenzhen Best Electronics Co.,Ltd All Rights Reserved.